LACTOSURE ECO DETAILS
- Fully indigenous Milk Analyser.
- World’s fastest Milk Analyser with measurement time of 27 seconds.
- Low maintenance cost and spares cost, timely service and ready availability of spares.
- Fast and accurate measurements in wide temperature range of 1˚C to 40˚C.
- Excellent repeatability.
- RS232 port for computer interfacing.
- Pest intrusion free design (The cabinet is fully sealed to avoid pest intrusion).
- Low maintenance design. (Ideal for Indian condition)
- Fully Indigenous manufacturing, enabling easy availability of spare parts and proper service and maintenance. All the materials including sensors, microprocessor integrated board, pump etc. are made in-house.
- Services and spares at affordable rates.
- Splash proof cabinet design. (Water or milk if spilled over the body, will not penetrate inside the cabinet. This keeps the electronics and its sensitive components safe and intact.)
- Cabinet made out of plastic and stainless steel is free from corrosion.
- High contrast blue/green backlight 4-line large character LCD display.
- Easily operational switch/buttons with 4 million operational capacities.
- External power supply adapter to avoid power line interference and easy maintenance.
- Low quantity milk sample required.
- Facility to backup/copy/save and restore calibration and settingsin system memory and also in computer.
- Two modes of cleaning, namely Daily and Weekly, for better overall performance and longevity of the equipment.
- Manual cleaning option using Plunger.
- Single sensor design with 27 seconds measurement time.
- Fully Indian: Fully indigenous make, selfsufficiency and ready availability of spares with very low spare cost. Costs of spares are extremely expensive for the imported machines.
- No need of changing Mother board (Main board) at the time of sensor replacement.
- Easy sensor replacement and calibration even at customer place.
- Special training: Hardware and software training for dealer, for self reliance and better profit margins
- Pest intrusion free design (The cabinet is fully sealed to avoid pest intrusion).If the cabinet have air holes, which the pest/insects will come inside the analyser and damage the silicon tube.
- All mechanical parts and cabinet made out of plastic, chromium plated brass and 304 grade stainless steel (even screws) is free from corrosion.
- External power supply adapter to avoid power line interference and easy maintenance.
- Splash proof cabinet design. (Water or milk if spilled over the body, will not penetrate inside the cabinet. This keeps the electronics and its sensitive components safe and intact.
- 3 level password protection for easily service at site.
- Facility to set CLR formulaaccording to State by state.
Direct Measuring Parameters |
FAT: 0.5% to 15 % with accuracy +/-0.1% |
Calculated Parameters |
CLR: 20 to 40 with accuracy +/-1 |
Resolution |
FAT, SNF, Lactose, Protein, Salts: 0.01% |
Repeatability |
FAT: +/-0.07% |
Milk temperature |
1 to 40°C |
Measurement Time |
27 Sec @30°C |
Sample Volume |
15ml |
Average speed |
120 samples/hr |
Number of Calibrations |
3 |
Operation time Limit |
Continuous |
Display |
High contrast 4 line, 20 Character with Blue/Green Backlight |
Operating Voltage |
12 Volts DC (140 – 250V AC for External Adapter ) |
Power input |
With external SMPS power supply. |
Average Power |
30W |
Peak Power |
120W |
Computer Interface |
RS232 |
Cleaning modes |
Daily, Weekly |
Cabinet |
ABS Front and rest 304 gradeStainless Steel |
Protection |
Splash proof design, All mechanical parts made of304 grade S.S, Brass and ABS. |
Environmental Conditions |
|
Accessories |
|